The delivery package contains the following items

1 Heat Sealer

1 Base Plate for 96 well PCR plates

1 Instruction Manual

1 Power lead

In order to avoid damage caused by condensation, the device may first be switched on
12 hours after setting up. This also applies to a change in location.

The Heat Sealer must stand completely stable on the work surface in a safe, dry location away
from the edge of the work surface.

Space requirement:

Width: 14 cm
Depth: 15 cm
Height: 33 cm

Mains connection:

An earthed wall socket for the Heat Sealer.

The mains voltage must agree with the information on the name plate.

The device must be connected to a grounded socket.

The delivery carton should be kept in order to be able to ship the device in case repairs are
necessary.