9 Technical data
Centrifuge 5424 / 5424 R — Operating manual
34
9
Technical data
Operating manual
—
Centrifuge 5424 / 5424 R
see
on p.
Fig.
Tab.
p.
9.1
Power supply
Centrifuge 5424
Centrifuge 5424 R
9.2
Ambient conditions
Mains power connection
230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz
Current consumption:
1.9 A (230 V)
3.8 A (120 V)
4.0 A (100 V)
Power consumption:
max. 250 W
EMC: Interference emission (radio
interference)
EN 61326 - category B
EMC: Noise immunity
EN 61326
Overvoltage category:
II
Fuses:
3.15 AT (230 V)
6.3 AT (120 V / 100 V)
Mains power connection
230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz
Current consumption:
1.6 A (230 V)
3.2 A (120 V)
3.4 A (100 V)
Power consumption:
max. 350 W
EMC: Interference emission (radio
interference)
EN 61326 - category B
EMC: Noise immunity
EN 61326
Overvoltage category:
II
Fuses:
3.15 AT (230 V)
6.3 AT (120 V / 100 V)
Environment:
For indoor use only.
Ambient temperature:
Centrifuge 5424: 2 to 40°C
Centrifuge 5424 R: 10 to 40°C
Max. relative humidity:
10 to 75 %, non-condensing humidity
Atmospheric pressure:
Use up to an altitude of 2000 m above MSL.
Degree of contamination:
2
EN