7 troubleshooting, 1 general errors, Troubleshooting 7.1 – Eppendorf Easypet 3 User Manual
Page 28: General errors
28
Troubleshooting
Easypet
®
3
English (EN)
7
Troubleshooting
7.1
General errors
Symptom/message
Cause
Remedy
Liquid drips out of the
pipette.
• Pipette adapter and/or
membrane filter inserted
incorrectly.
Remove the pipette adapter
and membrane filter and
reinsert them.
• Pipette not inserted far
enough.
Continue to carefully insert
the pipette.
• Pipette adapter damaged.
Replace pipette adapter.
• Pipette damaged.
Replace pipette.
• Seal damaged.
Replace seal.
• Liquid column too high.
• Speed too low.
Slowly press the aspirating
button down further.
Aspiration capacity
reduced.
• Membrane filter wetted.
Replace membrane filter.
No functional period
despite fully charged
battery.
• Battery is too old.
Replace battery.
Pipette loose.
• Pipette adapter damaged.
Replace pipette adapter.
Bubble formation in
the pipette during
liquid aspiration.
• Too high speed.
The aspirating button may
only be pressed lightly.
• Too high speed.
The aspirating button may
only be pressed lightly.
Rechargeable battery
will not charge.
• Rechargeable battery is
charged.
Disconnect the mains/power
supply device.
Only charge the
rechargeable battery if the
status display is blinking.