beautypg.com

7 troubleshooting, 1 general errors, Troubleshooting 7.1 – Eppendorf Easypet 3 User Manual

Page 28: General errors

background image

28

Troubleshooting
Easypet

®

3

English (EN)

7

Troubleshooting

7.1

General errors

Symptom/message

Cause

Remedy

Liquid drips out of the
pipette.

• Pipette adapter and/or

membrane filter inserted
incorrectly.

Remove the pipette adapter
and membrane filter and
reinsert them.

• Pipette not inserted far

enough.

Continue to carefully insert
the pipette.

• Pipette adapter damaged.

Replace pipette adapter.

• Pipette damaged.

Replace pipette.

• Seal damaged.

Replace seal.

• Liquid column too high.

• Speed too low.

Slowly press the aspirating
button down further.

Aspiration capacity
reduced.

• Membrane filter wetted.

Replace membrane filter.

No functional period
despite fully charged
battery.

• Battery is too old.

Replace battery.

Pipette loose.

• Pipette adapter damaged.

Replace pipette adapter.

Bubble formation in
the pipette during
liquid aspiration.

• Too high speed.

The aspirating button may
only be pressed lightly.

• Too high speed.

The aspirating button may
only be pressed lightly.

Rechargeable battery
will not charge.

• Rechargeable battery is

charged.

Disconnect the mains/power
supply device.

Only charge the
rechargeable battery if the
status display is blinking.