Page 20: 3english (en), Fig. 5-3: slow aspiration

background image

20

Operation
Easypet

®

3

English (EN)

5.4

Speed control

The liquid aspiration speed is regulated continuously by controlling how far the control
button is pressed.

Slowly aspirating or dispensing liquid

Type 3 and type 4 measuring pipette

Measuring pipette

• Adjusted to flow-out.

• Nominal volume is indicated by the tip of the pipette.

• Aspiration up to the zero line at the upper end of the pipette.

• Dispensing to any scale graduation on the scale or until the pipette is

completely empty.

Type 4 measuring pipette

• Adjusted to blow-out.

• Nominal volume is indicated by the tip of the pipette.

• Aspiration up to the zero line at the upper end of the pipette.

• Dispensing to any scale graduation on the scale or until the pipette is

completely empty.

• Dispensing the last drops using blow-out.

• Blow-out pipettes are only available in accuracy class B.

Abb. 5-3:Slow aspiration

Fig. 5-3:

Slow aspiration

To slowly aspirate or dispense liquid,
press the corresponding control button
lightly.