Warranty, Error codes – IKA EUROSTAR 100 digital User Manual
Page 10

10
Warranty
In accordance with IKA
®
warranty conditions, the
warranty period is 24 months. For claims under
the warranty please contact your local dealer. You
may also send the machine direct to our factory,
enclosing the delivery invoice and giving reasons
for the claim. You will be liable for freight costs.
The warranty does not cover worn out parts, nor
does it apply to faults resulting from improper use,
insufficient care or maintenance not carried out in
accordance with the instructions in this operating
manual.
Error codes
The fault is shown by an error code in the display (C) as following if the error occurs, e.g. Er 4.
Proceed as follows in such cases:
F
Switch the device switch (A) off.
F
Remove the stirrer tool and remove the instrument from the assembly.
F
Reduce the speed and switch on (instrument switch (A)) the device without the stirrer tool.
Error code Cause
Effect
Solution
Er 2
Motor current sensor fault
motor off
- Switch off the instrument
Er 3
Temperature inside instrument
is too high
motor off
- Switch off the instrument and allow it
to cool down
Er 4
Motor blockage or overload
motor off
- Switch off the instrument
- Decrease the load on the motor and
restart again
Er 8
Speed sensor fault or overload motor off
- Switch off the instrument
Er 21
Safety relay fault
motor off
- Switch off the instrument
If the actions described fails to resolve the fault or another error code is displayed then take one of the
following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.