IKA RET control-visc User Manual
Page 31

31
NAMUR commands
Function
Display
(additional)
IN_NAME
Title request
IN_PV_X X=1;2;3;4;5;7;80;90 Current value reading
IN_SOFTWARE
Software ID number, date, version request
IN_SP_X
X=1;2;3;4;7;8;12;
42;54;55;56;
Set target value reading
IN_TYPE
Lab device identification request
OUT_NAME name
Output of identification name. (Max. 6 characters; default: IKAMAG)
OUT_SP_12@n
Setting WD safety limit temperature with set value echo
OUT_SP_42@n
Setting WD safety limit speed with set value echo
OUT_SP_X n
X=1;2;4;7;
54;55;56
Setting of target value to n
OUT_WD1@m
Watchdog mode 1: if event WD1 should occur, the heating and stirring functions are
switched off and ER 2 is displayed. Set watchdog time to m (20 - 1,500) seconds,
with watchdog time echo. This command launches the watchdog function and must
be transmitted within the set watchdog time.
OUT_WD2@m
Watchdog mode 2: if event WD2 should occur, the speed target value is changed to
the WD safety speed limit and the temperature target value is changed to the WD
safety temperature limit value. The warning WD is displayed. The WD2 event can be
reset with OUT_WD2@0 - this also stops the watchdog function. Set watchdog time
to m (20 - 1,500) seconds, with watchdog time echo. This command launches the
watchdog function and must be transmitted within the set watchdog time.
RESET
Switches off the device function.
START_X X=1;2;4;5;7
X=80; 90
Switching on of device - (remote) function
Activation of the previous watchdog function. If no target value is entered within
60 seconds, the (old function) heating is switched off.
Remote
STOP_X X=1;2;4;5;7
X=80; 90
Switching off of device - (remote) function The variables set with OUT_SP_X remain
saved. Includes command RMP_STOP.
Remote