Technical Data
RCT basic safety control
HCT basic safety control
Device
Operating voltage range
Vac
220 - 230 ± 10%
Vac
115 ± 10%
Vac
100 ± 10%
Nominal voltage
Vac
230/ 50 Hz
Vac
115/ 60 Hz
Vac
100/ 60 Hz
Frequency
Hz
50 / 60
Power consumption (+10%) max. at 230 Vac
W
650
640
115 Vac
610
600
100 Vac
610
600
Display
digital
Permissible duration of operation
%
100
Permissible ambient temperature
°C
+5 to +40
Permissible relative humidity
%
80
Protection tzype acc. DIN EN 60529
IP 42
Protection class
I
Overvoltage category
II
Contamination level
2
Fuse
F1/F2
T6,3A/250V (at nominal voltage 230 V)
T10A/250V (at nominal voltage 115 V and 100 V)
Operation at a terrestrial altitude
m
max. 2000
Dimensions (B x T x H)
mm
165 x 275 x 85
Wight
kg
2,5
Motor
Speed range
rpm
50 - 1500
Power consumption
W
16
Setting resolution
rpm
10
Speed variation (no load, nominal voltage,
%
±2
at 1500 rpm, ambient temperature 25 °C)
Stirred quantity max. (H
2
O)
ltr
20
22