Error codes – IKA ROCKER 3D digital User Manual
Page 17

17
Repair
In case of repair the device has to be cleaned and free from any
materials which may constitute a health hazard.
For this you should use the downloaded print of the "Certifi cate
of compliance" form from the IKA
®
website www.ika.com.
If you require servicing, return the appliance in its original pa-
ckaging. Storage packaging is not suffi cient. Please also use sui-
table transport packaging.
Spare parts order
When ordering spare parts, please give:
- Machine type
- Manufacturing number, see type plate
- Item and designation of the spare part, see www.ika.com,
spare parts diagram and spare parts list
Error codes
IKA
®
Rocker 3D basic
In case of a fault, a beep is heard and the “Power” LED fl ashes.
Proceed as follows in such cases:
Switch the device off.
Perform corrective measures as described below.
Perform a restart.
Error
Cause
Effect
Solution
Motor blockage or overload
Motor off
- Switch the device off.
- Reduce the load on the motor and check all moving parts for
freedom of movement.
- Then perform a restart.
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the IKA
®
service department,
- Send the device for repair, including a short description of the fault.
IKA
®
Rocker 3D digital
The fault is shown by an error code in the display as following if the error occurs, e.g. Er 4.
Proceed as follows in such cases:
Switch the device off.
Perform corrective measures as described below.
Perform a restart.
Error
Cause
Effect
Solution
Er 3
Temperature inside device is too
high
Motor off
- Switch the device off.
- Allow the device to cool down.
Er 4
Motor blockage or overload
Motor off
- Switch the device off.
- Reduce the load on the motor and check all moving parts for
freedom of movement.
- Then perform a restart.
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the IKA
®
service department,