Operating modes A and B
Operating mode A
The device does not restart automatically when power is restored after a power outage.
Operating mode B
The device restarts automatically when power is restored after a power outage, using the set values, and displays the (Pr) symbol to
indicate that there has been a power outage within a certain period of time.
Switching between operating modes
To switch between operating modes, switch the device off, then keep the "Start/Stop" button (F) depressed and press the "Power"
button (L).
Adjusting the inclination angle of the shaking table
The inclination angle of the shaking table can be adjusted steplessly from 0 - 15°.
Caution! Do not remove the mat (1) and the cover (2) whilst the device is in operation!
Stop the device by pressing the "Start/Stop" button (F). Do not use your hand to stop the device.
Press the "Power" button (L) to switch the device off.
- To adjust the inclination angle of the shaking table, fi rst remove the mat (1), then the cover (2) that lies beneath it. To do this, press
the recess.
- Slacken the screw (3), using the 4 AF hexagon L-wrench to DIN 911 supplied; prevent the rotary knob from turning.
Caution! Do not fully undo the screw! Slackening the screw half a turn is quite suffi cient to allow adjustment of the inclination angle
of the shaking table.
- Turn the rotary knob to adjust the inclination angle of the shaking table to the desired angle (4).
- Tighten the screw (3) fi nger tight.
- Refi t the cover (2).
3
4
Use of accessories
2