Specifications, Specifications intellitherm, Model tgt70 – WIKA TGT70 User Manual
Page 8: Electronics

14047194.02 02/2013 GB/D/F/E
8
WIKA operating instructions model TGT70
GB
3. Specifications
Specifications
intelliTHERM
®
model TGT70
Measuring principle
Bourdon tube system
Nominal size in mm
63, 100
Filling medium
Xylol or silicone oil
Connection location
Lower mount (LM) (radial)
Version B: back mount (axial)
Instrument version
■
H
■
M
■
B
■
V
■
R
Instrument with capillary and rear surface mounting flange
Instrument with capillary and surface mounting bracket
Instrument with capillary, triangular bezel and mounting clamp
Instrument with capillary and panel mounting flange
Direct connection without capillary
Indication accuracy
Class 2, EN 13190
Capillary
Length in accordance with customer specifications (max. 10 m)
Ø 2 mm, stainless steel 1.4571, bending radius no less than 6 mm
Capillary entry
lower mount
Capillary mounting
Take care that the mounting is free from vibration
Case and bayonet ring
Stainless steel
Connection
plain, stainless steel 1.4571
Stem
Ø 8 mm, stainless steel 1.4571
Active sensor length
Depends on Ød and scale range
Dial
Plastic sticker, white with logo
Aluminium, white, black lettering
Pointer
Aluminium, black
Window
Laminated safety glass
Ingress protection
IP 65 per EN/IEC 60529
Electronics
Output signal
■
Voltage output
With U
s
= DC 5 V, ratiometric: 0.5 … 4.5 V
With U
s
= DC 12 … 32 V, non ratiometric: 0.5 … 4.5 V
■
Current output
4 … 20 mA, 2-wire
Power supply (U
s
)
DC 5 V / DC 12 … 32 V
Electromagnetic compatibility EMC directive 2004/108/EC, EN 61326 emission (group 1,
class B) and interference immunity (industrial application)
3. Specifications