3. Specifications

Specifications

intelliTHERM

®

model TGT70

Measuring principle

Bourdon tube system

Nominal size in mm

63, 100

Filling medium

Xylol or silicone oil

Connection location

Lower mount (LM) (radial)

Version B: back mount (axial)

Instrument version

H

M

B

V

R

Instrument with capillary and rear surface mounting flange

Instrument with capillary and surface mounting bracket

Instrument with capillary, triangular bezel and mounting clamp

Instrument with capillary and panel mounting flange

Direct connection without capillary

Indication accuracy

Class 2, EN 13190

Capillary

Length in accordance with customer specifications (max. 10 m)

Ø 2 mm, stainless steel 1.4571, bending radius no less than 6 mm

Capillary entry

lower mount

Capillary mounting

Take care that the mounting is free from vibration

Case and bayonet ring

Stainless steel

Connection

plain, stainless steel 1.4571

Stem

Ø 8 mm, stainless steel 1.4571

Active sensor length

Depends on Ød and scale range

Dial

Plastic sticker, white with logo

Aluminium, white, black lettering

Pointer

Aluminium, black

Window

Laminated safety glass

Ingress protection

IP 65 per EN/IEC 60529

Electronics

Output signal

Voltage output

With U

s

= DC 5 V, ratiometric: 0.5 … 4.5 V

With U

s

= DC 12 … 32 V, non ratiometric: 0.5 … 4.5 V

Current output

4 … 20 mA, 2-wire

Power supply (U

s

)

DC 5 V / DC 12 … 32 V

Electromagnetic compatibility EMC directive 2004/108/EC, EN 61326 emission (group 1,

class B) and interference immunity (industrial application)

3. Specifications