beautypg.com

Faults – WIKA TR10-J User Manual

Page 17

background image

WIKA operating instructions resistance thermometers, thermocouples

11208287.06 10/2013 GB/D/F/E

17

GB

8. Faults

7.3 Calibration, Recalibration

It is recommended that the measuring insert is recalibrated at regular intervals (resistance

thermometers: approx. 24 months, thermocouples: approx. 12 months). This period can be

reduced, depending on the particular application. The calibration can be carried out by the

manufacturer, as well as on site by qualified technical staff with calibration instruments.

8. Faults

Faults

Causes

Measures

No signal/line breakage

Mechanical load too high or

overtemperature

Replace probe or measuring insert

with a suitable design

Erroneous measured values Sensor drift caused by

overtemperature

Replace probe or measuring insert

with a suitable design

Sensor drift caused by chemical attack Use a design with thermowell

Erroneous measured values

(too low)

Entry of moisture into cable or

measuring insert

Replace probe or measuring insert

with a suitable design

Erroneous measured values

and response times too long

Wrong mounting geometry, for example

mounting depth too deep or heat

dissipation too high

The temperature-sensitive area of the

sensor must be inside the medium,

and surfaces must be isolated.

Deposits on the sensor or thermowell Remove deposits

Erroneous measured values

(of thermocouples)

Parasitic voltages (thermal voltages,

galvanic voltage) or wrong equalisation

line

Use a suitable equalisation line

Indication of the measured

value jumps

Cable break in connecting cable or

loose contact caused by mechanical

overload

Replace probe or measuring insert

with a suitable design, for example

equipped with a strain relief or a

thicker conductor cross-section

Corrosion

Composition of the medium not

as expected or modified or wrong

thermowell material selected

Analyse medium and then select a

more-suitable material or replace

thermowell regularly

Signal interference

Stray currents caused by electric fields

or earth loops

Use of screened connecting cables,

increase in the distance to motors and

power lines

Earth circuits

Elimination of potentials, use of

galvanically isolated transmitter supply

isolators or transmitters

CAUTION!

If deficiencies cannot be eliminated by means of the measures listed above, shut

down the instrument immediately, and ensure that pressure and/or signal are no

longer present, and secure the instrument from being put back into operation

inadvertently.

In this case, contact the manufacturer.

If a return is needed, please follow the instructions see chapter 9.2 “Return”.

This manual is related to the following products: