WIKA Operating instructions pressure gauges Model 7 per ATEX

7

GB

2080284.06 10/2013 GB/D/F

Movement:

Stainless steel

Dial and pointer: Aluminium

Case, bezel ring: Stainless steel

(Model 73x.31: with blow-out back)

Window:

Laminated safety glass

Installation

„

Nominal position per EN 837-3 / 9.6.6 Figure 7: 90° (

)

„

Process connection: lower mount (LM)

„

In order to ensure that, with Model 73x.31 pressure can be safely and

reliably vented through the case back, a distance of at least 25 mm must be
left free behind the case!

„

In order to avoid any additional heating, the instruments must not be

exposed to direct solar irradiation while in operation!

„

With filled versions the ventilating valve at the top of the case must be

opened prior to commissioning!

Permissible vibration load at the mounting point

„

The instruments should always be installed in locations free from vibration.

„

If necessary, it is possible to isolate the instrument from the mounting point

by installing a flexible connection line between the measuring point and the
pressure gauge and mounting the instrument on a suitable bracket.

„

If this is not possible, the following limit values must not be exceeded:

Dry gauges:

Frequency range < 150 Hz

(Model 732)

Acceleration < 0.7 g (7 m/s

2

)

Liquid-filled gauges: Frequency range < 150 Hz
(Model 733)

Acceleration < 4 g (40 m/s

2

)

The liquid filling must be checked on a regular basis.
The liquid level must not drop below 75 % of the gauge diameter.

3. Technical data and intended use

This manual is related to the following products: