Commissioning, Maintenance and cleaning, Appendix 1: declaration of conformity – WIKA 111.11 User Manual
Page 3
Permissible vibration load at the installation site
■
The instruments should always be installed in locations free from vibration.
■
If necessary, it is possible to isolate the instrument from the mounting point by installing
a flexible connection line between the measuring point and the pressure gauge and
mounting the instrument on a suitable bracket.
■
If this is not possible, the following limit values must not be exceeded:
Dry gauges:
Frequency range < 150 Hz
(Model 111)
Acceleration < 0.7 g (7 m/s
2
)
Liquid-filled gauges: Frequency range < 150 Hz
(Model 113, 213)
Acceleration < 4 g (40 m/s
2
)
The liquid filling must be checked on a regular basis.
The liquid level must not drop below 75 % of the gauge diameter.
4. Commissioning
During the commissioning process pressure surges must be avoided at all costs. Open the
shut-off valves slowly.
5. Maintenance and cleaning
■
The instruments are maintenance-free.
■
The indicator should be checked once or twice every year. For this the instrument must
be disconnected from the process to check with a pressure testing device.
■
Clean the pressure gauge with a moist cloth.
■
Repairs must only be carried out by the manufacturer or appropriately qualified skilled
personnel.
■
When dismounting, close the compensating valve (if available).
WARNING!
Residual media in dismounted pressure gauges can result in a risk to persons,
the environment and equipment.
Take sufficient precautionary measures.
Appendix 1: Declaration of conformity