Sealing in the

thread 

Spanner flats

Sealing face 

Correct sealing of pressure gauge connections with parallel threads  must be made using

suitable flat gaskets, sealing rings or WIKA profile sealings. The sealing of tapered threads

(e.g. NPT threads) is made by providing the thread  with additional sealing material such

as, for example, PTFE tape (EN 837-2).

The torque depends on the seal used. Connecting the gauge using a clamp socket or a

union nut is recommended, so that it is easier to orientate the gauge correctly.
When a blow-out device is fitted to a pressure gauge, it must be protected against being

blocked by debris and dirt.
After mounting, set the compensating valve (if available) from CLOSE

to OPEN.

Temperature effect
When the temperature of the measuring system deviates from the reference temperature

(+20 °C):

max. ±0.4%/10 K of full scale value

Ingress protection per EN 60529 / IEC 529

Model 111.xx:

IP 42

Models 113.53, 213.40 and 213.53: IP 65

Permissible temperatures
Ambient:

Model 111.xx:

-40 ... +60 °C

Models 113.53, 213.40 and 213.53: -20 ... +60 °C

Medium: max. +60 °C

Attention! With gaseous substances, the temperature may increase as a result of

compression warming. In these cases it may be necessary to throttle the rate of change

of pressure or reduce the permissible medium temperature.
The effective maximum surface temperature is not only dependant upon these instruments,

but mainly on the respective medium temperature!

Materials
Wetted parts: Cu-alloy
Movement:

Cu-alloy

Dial:

NS 40, 50, 63: Plastic

NS 80, 100, 160: Aluminium

Pointer:

Plastic (NS 160: Aluminium)

Case:

Models 111.10, 111.12, 111.16: Plastic

Model 111.11:

Steel

Models 113.53, 213.53:

Stainless steel

Model 213.40:

Forged brass

Window:

Plastic (NS 160: Instrument glass)

Installation

Nominal position per EN 837-1 / 9.6.7 Figure 9: 90° ( ⊥ )

Process connection lower mount (LM) or back mount (BM)

In order to avoid any additional heating, the instruments must not be exposed to direct
solar irradiation while in operation!

Pressure gauges must be earthed via the process connection!

Permissible ambient and operating temperatures
When mounting the pressure gauge it must be ensured that, taking into consideration the

influence of convection and heat radiation, no deviation above or below the permissible

ambient and medium temperatures can occur. The influence of temperature on the

indication accuracy must be observed.