beautypg.com

Gb 3. specifications 3. specifications – WIKA LH-10 User Manual

Page 8

background image

8

WIKA operating instructions submersible pressure transmitter, model LH-10

14075011.01 08/2013 GB/D/F/E

GB

3. Specifications

3. Specifications

Measuring ranges

Relative pressure

bar

Measuring range

0 ... 0.1

0 ... 0.16

0 ... 0.25

0 ... 0.4

0 ... 0.6

0 ... 1

0 ... 1.6

Overpressure limit

1

1.5

2

2

3

5

8

Burst pressure

2

2

2.4

2.4

4

6

10

Measuring range

0 ... 2.5

0 ... 4

0 ... 6

0 ... 10

0 ... 16

0 ... 25

Overpressure limit

8

10

10

10

16

25

Burst pressure

10

10

10

10

16

25

inWC

Measuring range

0 ... 50

0 ... 100

0 ... 150

0 ... 250

Overpressure limit

750

750

750

1,100

Burst pressure

950

950

950

1,600

psi

Measuring range

0 ... 5

0 ... 10

0 ... 15

0 ... 25

0 ... 50

0 ... 100

0 ... 150

Overpressure limit

30

45

70

120

150

150

150

Burst pressure

35

60

90

180

150

150

150

Measuring range

0 ... 160

0 ... 200

0 ... 300

Overpressure limit

160

200

300

Burst pressure

160

200

300

mH

2

O Measuring range

0 ... 1

0 ... 1.6

0 ... 2.5

0 ... 4

0 ... 6

0 ... 10

0 ... 16

Overpressure limit

10

15

20

20

30

50

80

Burst pressure

20

20

24

24

40

60

100

Measuring range

0 ... 25

0 ... 40

0 .. 60

0 ... 100

0 ... 160

0 ... 250

Overpressure limit

80

100

100

100

160

250

Burst pressure

100

100

100

100

160

250

When choosing the FEP cable, measuring ranges up to and including 0 ...10 bar, 0 ... 150 psi and 0 ... 100 mH

2

O are

available.

The given measuring ranges are also available in mbar, kPa and MPa.