beautypg.com

Dismounting, return and disposal, Faults / 9. dismounting, return and disposal – WIKA LH-10 User Manual

Page 14

background image

14

WIKA operating instructions submersible pressure transmitter, model LH-10

14075011.01 08/2013 GB/D/F/E

GB

Faults

Causes

Measures

Signal span too small/drops

Mechanical overload caused by overpres-

sure

Replace instrument; if it fails repeatedly, contact

the manufacturer.

Diaphragm damaged, e.g. due to impacts,

abrasive/aggressive medium; corrosion at

diaphragm; transmission medium missing

Replace instrument; if it fails repeatedly, contact

the manufacturer.

Signal span varies/inaccurate

Operating temperature too high/low

Operate the instrument in the permissible

temperature range

Deviating zero point signal

Operating temperature too high/low

Operate the instrument in the permissible

temperature range

Diaphragm damaged, e.g. due to impacts,

abrasive/aggressive medium; corrosion at

diaphragm; transmission medium missing

Replace instrument; if it fails repeatedly, contact

the manufacturer.

If complaint is unjustified, we will charge you the complaint processing fees.

CAUTION!

If faults cannot be eliminated by means of the measures listed above, shut down the submersible pressure

transmitter immediately, and ensure that pressure is no longer present, and secure the instrument from

being put back into operation inadvertently. In this case, contact the manufacturer. If a return is needed,

please follow the instructions given in chapter 9.2 “Return”.

9. Dismounting, return and disposal

WARNING!

Residual media in the dismounted instrument can result in a risk to persons, the environment and

equipment. Take sufficient precautionary measures.

9.1 Disassembly

WARNING!

Risk of burns!

Let the instrument cool down sufficiently before dismounting it!

During dismounting there is a risk of residual, dangerously hot media.

8. Faults / 9. Dismounting, return and disposal