Dismounting, return and disposal, Faults / 9. dismounting, return and disposal – WIKA LH-10 User Manual
Page 14
14
WIKA operating instructions submersible pressure transmitter, model LH-10
14075011.01 08/2013 GB/D/F/E
GB
Faults
Causes
Measures
Signal span too small/drops
Mechanical overload caused by overpres-
sure
Replace instrument; if it fails repeatedly, contact
the manufacturer.
Diaphragm damaged, e.g. due to impacts,
abrasive/aggressive medium; corrosion at
diaphragm; transmission medium missing
Replace instrument; if it fails repeatedly, contact
the manufacturer.
Signal span varies/inaccurate
Operating temperature too high/low
Operate the instrument in the permissible
temperature range
Deviating zero point signal
Operating temperature too high/low
Operate the instrument in the permissible
temperature range
Diaphragm damaged, e.g. due to impacts,
abrasive/aggressive medium; corrosion at
diaphragm; transmission medium missing
Replace instrument; if it fails repeatedly, contact
the manufacturer.
If complaint is unjustified, we will charge you the complaint processing fees.
CAUTION!
If faults cannot be eliminated by means of the measures listed above, shut down the submersible pressure
transmitter immediately, and ensure that pressure is no longer present, and secure the instrument from
being put back into operation inadvertently. In this case, contact the manufacturer. If a return is needed,
please follow the instructions given in chapter 9.2 “Return”.
9. Dismounting, return and disposal
WARNING!
Residual media in the dismounted instrument can result in a risk to persons, the environment and
equipment. Take sufficient precautionary measures.
9.1 Disassembly
WARNING!
Risk of burns!
Let the instrument cool down sufficiently before dismounting it!
During dismounting there is a risk of residual, dangerously hot media.
8. Faults / 9. Dismounting, return and disposal