Fault Remedy
If any fault occurs, such fault may be caused by a total battery discharge or too
high or too low a temperature. Therefore, it is recommended to restart the
charging after lapse of some waiting time. Such charging may be accomplished
either by:
•
pulling the plug charger (6) out of the 220V mains outlet and reconnecting
it; or
•
taking the DOCTER
aspherilux midi LED out and inserting it again;
or
•
switching the DOCTER
aspherilux midi LED on and off again.
Please note that the battery will not be charged as long as the LED flashlight is
switched on. There will be no charging unless the DOCTER-aspherilux LED
spotlight is switched off.
The contact surfaces (4) on the bottom side of the DOCTER
aspherilux midi LED
and the contact pins at the charging (7) station should be kept clean (to be
cleaned with spirit) and protected from damage.
Battery Warning
During operation, the battery voltage is measured permanently. If such
measurement detects that the battery voltage is nearing the programmed warning
voltage, then two short flashes will be released as a visual warning signal. This
signal is sent off in switched-on condition once only. Upon re-start and after a
short pause of approximately one second, two short flash signals will recall the
warning condition. Depending on the temperature and frequency of use, a 10 to 30
minutes operation back-up remains available.
Automatic Switch-Off and Emergency Light
When the final discharge voltage of the cells is reached, a permanent flashing will
be initiated in order to avoid total discharge. Then, the LED will always be
switched shortly with longer pauses. Thus, an emergency light is available. To
avoid permanent damage to the batteries, operation in this condition should be
limited to a minimum.
Environmental Influences
When leaving the factory, the DOCTER
aspherilux midi LED is waterproof.