Troubleshooting – BrandTech Scientific Transferpette S User Manual
Page 46

Troubleshooting
Problem
Possible cause
Corrective action
Tip dripping
(instrument leaks)
Unsuitable tip
Only use high-quality
tips
Tip not seated tightly
Press tip on firmly
The instrument does
not aspirate or as-
pirates too little; the
discharged volume
is too low.
Seal contaminated
Clean seal
The seal or cone is
damaged
Replace seal or shaft
The piston is contami-
nated or damaged
Clean or replace piston
Aspiration is
too slow
Shaft clogged
Clean shaft
The filter in the 5 ml
and 10 ml models is
contaminated
Change the filter
Discharged volume
is too large
Pipetting button
pressed too far into
the blow-out position
before sample uptake
Operate properly.
See 'Pipetting',
page 32.
Piston is difficult
to move
The piston is contami-
nated or needs oil
Clean and oil the piston
46