Accuracy table, English 37, Transferpette – BrandTech Scientific Transferpette S User Manual
Page 37: S, adjustable volume, 20 °c ex

English
37
Accuracy Table
Transferpette
®
S, adjustable volume
!
20 °C
Ex
Final test values related to the nominal capacity (maximum volume)
indicated on the instrument, and the indicated volume steps obtained when
instrument and distilled water are equilibrated at ambient temperature
(20 °C/68 °F) and with smooth operation. According to DIN EN ISO 8655.
* A = Accuracy, CV = Coefficient of Variation
Volume range
µl
Volume step
µl
A*
≤ ± %
CV*
≤ %
Increment
µl
Recommended
type of tip, µl
0.1 - 1
1
2
1.2
0.001
0.1 - 20
0.5
4
2.4
0.1
20
12
0.1 - 2.5
2.5
1.4
0.7
0.002
0.5 - 20
1.25
2.5
1.5
0.25
12
6
0.5 - 10
10
1
0.5
0.01
0.5 - 20
5
1.6
1
1
7
4
2 - 20
20
0.8
0.4
0.02
2 - 200
10
1.2
0.7
2
5
2
10 - 100
100
0.6
0.2
0.1
2 - 200
50
0.8
0.4
10
3
1
20 - 200
200
0.6
0.2
0.2
2 - 200
100
0.8
0.3
20
3
0.6
100 - 1000
1000
0.6
0.2
1
50 - 1000
500
0.8
0.3
100
3
0.6
500 - 5000
5000
0.6
0.2
5
500 - 5000
2500
0.8
0.3
500
3
0.6
1000 - 10000
10000
0.6
0.2
10
1000 - 10000
5000
0.8
0.3
1000
3
0.6