Accuracy table, Transferpette, S, fixed volume – BrandTech Scientific Transferpette S User Manual
Page 38: 20 °c ex

Transferpette
®
S, Fixed volume
Volume range
µl
A*
≤ ± %
CV*
≤ %
Recommended
type of tip, µl
10
1
0.5
0.5 - 20
20
0.8
0.4
2 - 200
25
0.8
0.4
2 - 200
50
0.8
0.4
2 - 200
100
0.6
0.2
2 - 200
200
0.6
0.2
2 - 200
500
0.6
0.2
50 - 1000
1000
0.6
0.2
50 - 1000
Accuracy Table
38
* A = Accuracy, CV = Coefficient of Variation
!
20 °C
Ex
Final test values related to the nominal capacity (maximum volume)
indicated on the instrument obtained when instrument and distilled
water are equilibrated at ambient temperature (20 °C/68 °F) and with
smooth operation. According to DIN EN ISO 8655.