II. Routine Maintenance

1. Always rinse the cell and electrodes with clean water after each use.

2. If the electrodes develop scaling or become dirty, clean the cell by

submerging the probe end in a 1:1 solution of Lime-A-Way® and water for 5

minutes. Then rinse thoroughly with clean water.

3. Do not drop, throw or otherwise strike the pen. This voids the warranty.

4. Do not store the pen in a location where the ambient temperatures exceed its

specified Operating/Storage Temperature limits.

SPECIFICATIONS

MYRON L

®

COMPANY

2450 Impala Drive • Carlsbad, Ca 92010-7226 • Phone: +1-760-438-2021

E-Mail: Customer Service – [email protected] • Technical Support – [email protected]

Website: www.myronl.com

MADE IN USA

©2014 MYRON L

®

COMPANY

App Requirements

Apple iOS 6 or iOS 7. Optimized for iOS 7.

Measurement Range:

Conductivity & TDS: 1-10,000 µS or ppm; Salinity 0.0010-

10.000 ppt; Temperature: 0°-71°C / 32°-160°F

Accuracy (After Wet Calibration):

± 1% of reading; Temperature ± 0.1ºC / ± 0.1ºF

Resolution:

Conductivity and TDS: 0.1 for 1.0 - 99.9 µS or ppm

1 for 100 - 10.000 µS or ppm

Salinity: 0.0001 for 0.0010 - 0.0999 ppt

0.001 for 0.100 - 10.000 ppt

Temperature: 0.1 ºC or ºF

Repeatability:

< 1000 µS or ppm ± 1 Count

≥ 1000 µS or ppm ± 0.3% of reading

Time to Reading Stabilization:

10 - 20 seconds

Active Mode Power Consumption: Active Mode 30 - 140 mA, Sleep Mode 2 µA
Temperature Compensation:

Automatic to 25ºC

Physical Dimensions:

17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D

Weight:

55 g or 1.94 oz

Case Material:

Anodized Aircraft Aluminum with Protective Coating

Battery Type:

N type, Alkaline 1.5 V

Calibration Standard Solution:

KCl: 1800 µS for COND; 442™: 3000 ppm (2027 ppm NaCl)

for TDS and Salinity.

Operating/Storage Temperature:

0 - 55ºC / 32 - 131ºF

Water Resistance:

IP67 and NEMA 6

EN61236-1: 2006 - Annex A:

2008; 1999/5/EC Annex III

Electrostatic discharge to the PTBT1 may

cause it to spontaneously power on. If this occurs,

the PTBT1 will power off after several seconds

FCC ID: T7VPAN17

The Bluetooth transceiver device meets the requirements for

modular transmitter approval as detailed in FCC public Notice

DA00-1407.

Canada (IC), license:

IC: 216Q-PAN17

The Bluetooth transceiver device meets the requirements for

modular transmitter approval as detailed in RSS-GEN.

ORIGINAL

EDITED FOR USE

Minimum height 5 mm

QUICK REFERENCE INSTRUCTIONS

1. Turn on ULTRAPEN App: Tap ULTRAPEN icon on mobile device’s home screen.

2. Turn on PTBT1: Press and release push button on pen.

3. LED flashes rapidly: Dip pen in sample solution so cell is totally submerged.

4. LED flashes slowly: Swirl pen to remove bubbles, keep cell submerged. Avoid contact with

sides/bottom of container,.

5. LED turns solid ON then turns OFF: Measurement is complete and reading is displayed.

Feature Navigation Bar

This button always returns the App to the Measurement Screen.

This button displays a list of saved measurements.

This button places the App in Calibration mode.

This button opens the App’s Bluetooth Connect / Disconnect screen.

This button opens the App’s preferences and settings mode.

An internet link to full Operations Manual (Help button) is found here.

ACCESSORIES

The ULTRAPEN™ PTBT1 uses the following solutions for wet calibration.

Order MODEL#s: KCL-1800, 442-3000 (2027 ppm NaCl)

MYRON L

®

WARRANTY

Excepting the battery, the Myron L

®

Company PTBT1 Conductivity/TDS/Salinity ULTRAPEN is

warranted to be free from manufacture defect for 1 year.