II. Routine Maintenance
1. Always rinse the cell and electrodes with clean water after each use.
2. If the electrodes develop scaling or become dirty, clean the cell by
submerging the probe end in a 1:1 solution of Lime-A-Way® and water for 5
minutes. Then rinse thoroughly with clean water.
3. Do not drop, throw or otherwise strike the pen. This voids the warranty.
4. Do not store the pen in a location where the ambient temperatures exceed its
specified Operating/Storage Temperature limits.
SPECIFICATIONS
MYRON L
®
COMPANY
2450 Impala Drive • Carlsbad, Ca 92010-7226 • Phone: +1-760-438-2021
E-Mail: Customer Service – [email protected] • Technical Support – [email protected]
Website: www.myronl.com
MADE IN USA
©2014 MYRON L
®
COMPANY
App Requirements
Apple iOS 6 or iOS 7. Optimized for iOS 7.
Measurement Range:
Conductivity & TDS: 1-10,000 µS or ppm; Salinity 0.0010-
10.000 ppt; Temperature: 0°-71°C / 32°-160°F
Accuracy (After Wet Calibration):
± 1% of reading; Temperature ± 0.1ºC / ± 0.1ºF
Resolution:
Conductivity and TDS: 0.1 for 1.0 - 99.9 µS or ppm
1 for 100 - 10.000 µS or ppm
Salinity: 0.0001 for 0.0010 - 0.0999 ppt
0.001 for 0.100 - 10.000 ppt
Temperature: 0.1 ºC or ºF
Repeatability:
< 1000 µS or ppm ± 1 Count
≥ 1000 µS or ppm ± 0.3% of reading
Time to Reading Stabilization:
10 - 20 seconds
Active Mode Power Consumption: Active Mode 30 - 140 mA, Sleep Mode 2 µA
Temperature Compensation:
Automatic to 25ºC
Physical Dimensions:
17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D
Weight:
55 g or 1.94 oz
Case Material:
Anodized Aircraft Aluminum with Protective Coating
Battery Type:
N type, Alkaline 1.5 V
Calibration Standard Solution:
KCl: 1800 µS for COND; 442™: 3000 ppm (2027 ppm NaCl)
for TDS and Salinity.
Operating/Storage Temperature:
0 - 55ºC / 32 - 131ºF
Water Resistance:
IP67 and NEMA 6
EN61236-1: 2006 - Annex A:
2008; 1999/5/EC Annex III
Electrostatic discharge to the PTBT1 may
cause it to spontaneously power on. If this occurs,
the PTBT1 will power off after several seconds
FCC ID: T7VPAN17
The Bluetooth transceiver device meets the requirements for
modular transmitter approval as detailed in FCC public Notice
DA00-1407.
Canada (IC), license:
IC: 216Q-PAN17
The Bluetooth transceiver device meets the requirements for
modular transmitter approval as detailed in RSS-GEN.
ORIGINAL
EDITED FOR USE
Minimum height 5 mm
QUICK REFERENCE INSTRUCTIONS
1. Turn on ULTRAPEN App: Tap ULTRAPEN icon on mobile device’s home screen.
2. Turn on PTBT1: Press and release push button on pen.
3. LED flashes rapidly: Dip pen in sample solution so cell is totally submerged.
4. LED flashes slowly: Swirl pen to remove bubbles, keep cell submerged. Avoid contact with
sides/bottom of container,.
5. LED turns solid ON then turns OFF: Measurement is complete and reading is displayed.
Feature Navigation Bar
This button always returns the App to the Measurement Screen.
This button displays a list of saved measurements.
This button places the App in Calibration mode.
This button opens the App’s Bluetooth Connect / Disconnect screen.
This button opens the App’s preferences and settings mode.
An internet link to full Operations Manual (Help button) is found here.
ACCESSORIES
The ULTRAPEN™ PTBT1 uses the following solutions for wet calibration.
Order MODEL#s: KCL-1800, 442-3000 (2027 ppm NaCl)
MYRON L
®
WARRANTY
Excepting the battery, the Myron L
®
Company PTBT1 Conductivity/TDS/Salinity ULTRAPEN is
warranted to be free from manufacture defect for 1 year.