A. Wet Calibration

Use a calibration solution specified for the measurement mode (see the Specification

table). Calibrating TDS calibrates Salinity for the same value and vice versa.

1. Pour Myron L

®

Company calibration solution into a clean container.

2. Rinse the cell 3 times by swirling it around in fresh calibration solution.

3. Open the PTBTx App.

4. Tap the CAL button in the Feature Navigation Bar. The App

screen will appear and say, “Paired ULTRAPEN is Offline”.

5. Grasp the pen near the pen cap to avoid sample contamination.

6. Remove pen from the rinse solution, then empty and refill the container with fresh

solution.

7. Press and release the pen’s push button. The Calibration screen will display, “Press

and hold the button on the ULTRAPEN to enter CAL mode.

8. Press and Hold the pen’s button. The LED will go to solid ON and the App’s Calibration

screen will appear.

9. Tap the green CAL button on the Calibration screen.

10. When the pen’s LED flashes rapidly, IMMEDIATELY dip

the pen in calibration solution so that the cell is completely submerged.

The Calibration screen will display. “Calibrating Conductivity ...”

11. When the LED flashes slowly, swirl the pen around to remove bubbles, keeping

the cell submerged. Keep pen at least 1 inch (2½ cm) away from sides/bottom of

container.

12. When the LED light goes solid ON, the Calibration screen will display

“CAL SAVED” along with the before and after calibration values.

B. Factory Calibration

If you do not have the proper calibration solution or wish erase the stored wet calibration

constants, use the FAC CAL function

1. Open the PTBTx App.

2. Tap the CAL button in the Feature Navigation Bar. The App screen will say, “Paired

ULTRAPEN is Offline.”

3. Press and release the push button. The screen will display, “Press and hold the

button on the ULTRAPEN to enter CAL mode”.

4. Press and Hold the ULTRAPEN’s button. The LED will go to solid ON and the App’s

Calibration screen will appear.

5. Tap the red FAC CAL button. A warning dialogue box

will appear.

6. Tap the OK button in the dialogue box. The App’s Calibrating screen will display,

“FAC CAL Reset”.

MAINTENANCE

I. Battery Replacement

The PTBT1 App display has an indicator that depicts the pen battery’s charge level.

When the charge level falls below 25%, the indicator will flash red. imediately replace

the battery with a new N type battery.

1. In a CLEAN DRY place unscrew the pen cap in a counter-clockwise direction.

2. Slide the cap and battery housing out of the pen.

3. Remove the depleted battery from its housing.

4. Insert a new battery into the battery housing oriented with the negative end touching

the spring.

5. Align the groove along the battery housing with the guide bump inside the pen case

and slide the battery housing back in.

6. Screw the pen cap back on in a clockwise direction.

... Continued from page 3

Do not over tighten.