Replacement ph sensors, Porta pak, Ordering – Myron L M6/PH User Manual
Page 6: Warranty & service

CALIBRATING THE RE-10
1. Calibrate your meter (without RE-10) using Standard Solution
number KCL-7000 (EP1 1/pH) or 442-3000 (M6/pH).
2. Fill cell cup with 442-30,000 high conductivity Standard Solution.
3. Insert RE-10 and press the button. Multiply the reading by 10
and compare it with the value on the Standard Solution label. If
they are not the same the RE-10 must be recalibrated.
NOTE: DO NOT adjust the Conductivity Calibrated Control.
4. Adjust the white insert of the Extender as follows: If the reading
is too high—push or tap inward. If reading is too low—twist or
pull outward with pliers.
NOTE: The Range Extender is factory calibrated to a particular meter.
It should be recalibrated if it is to be used with another meter.
REPLACEMENT pH SENSORS
Order Model RPY for either EP11/pH or M6/pH meters. When
ordering, be sure to include the model and serial number of your
meter to ensure receiving the proper type. Complete installation
instructions are provided with each replacement sensor.
PORTA PAK
Carrying Case for use with all
Myron L
DS and pDS
meters is
foam-lined and molded of sturdy
ABS plastic.
ORDERING
To order accessories contact your nearest stocking distributor, or
the Myron L Company.
WARRANTY & SERVICE
Myron L pDS meters, excluding the pH sensor, have a
two
-
year warranty. The pH sensor has a 6 month limited warranty.
If an instrument fails to operate properly, check the batteries,
calibrations and cleanliness of the pH sensor bulb. lf it still fails to
properly function, return it prepaid to the Myron L Company.
If, in the opinion of the factory, failure was due to materials or
workmanship, repair or replacement will be made without charge.
A reasonable service charge will be made for diagnosis
or
repairs
due
to normal
wear, abuse
or
tampering. This
warranty is
limited tothe repair or replacement of the Myron L
pDS meter
only. The Myron L Company assumes no other
responsibility or
liability.
2450 Impala Drive
Carlsbad, CA 920
1
0-
7226
USA
Tel 760-438-2021
Fax 800-869-7668 / 760-931-9189
www.myronl.com
pH Conductivity Instrumentation
Accuracy • Reliability • Simplicity
© Myron L Company 2001
(W)
Printed in U.S.A.
Specifications subject to change without notice