beautypg.com

Replacement ph sensors, Porta pak, Ordering – Myron L M6/PH User Manual

Page 6: Warranty & service

background image

CALIBRATING THE RE-10
1. Calibrate your meter (without RE-10) using Standard Solution

number KCL-7000 (EP1 1/pH) or 442-3000 (M6/pH).

2. Fill cell cup with 442-30,000 high conductivity Standard Solution.
3. Insert RE-10 and press the button. Multiply the reading by 10

and compare it with the value on the Standard Solution label. If
they are not the same the RE-10 must be recalibrated.

NOTE: DO NOT adjust the Conductivity Calibrated Control.
4. Adjust the white insert of the Extender as follows: If the reading

is too high—push or tap inward. If reading is too low—twist or
pull outward with pliers.

NOTE: The Range Extender is factory calibrated to a particular meter.
It should be recalibrated if it is to be used with another meter.

REPLACEMENT pH SENSORS

Order Model RPY for either EP11/pH or M6/pH meters. When

ordering, be sure to include the model and serial number of your
meter to ensure receiving the proper type. Complete installation
instructions are provided with each replacement sensor.

PORTA PAK

Carrying Case for use with all

Myron L

DS and pDS

meters is

foam-lined and molded of sturdy
ABS plastic.

ORDERING

To order accessories contact your nearest stocking distributor, or

the Myron L Company.

WARRANTY & SERVICE

Myron L pDS meters, excluding the pH sensor, have a

two

-

year warranty. The pH sensor has a 6 month limited warranty.
If an instrument fails to operate properly, check the batteries,
calibrations and cleanliness of the pH sensor bulb. lf it still fails to
properly function, return it prepaid to the Myron L Company.

If, in the opinion of the factory, failure was due to materials or

workmanship, repair or replacement will be made without charge.
A reasonable service charge will be made for diagnosis

or

repairs

due

to normal

wear, abuse

or

tampering. This

warranty is

limited tothe repair or replacement of the Myron L

pDS meter

only. The Myron L Company assumes no other

responsibility or

liability.

2450 Impala Drive

Carlsbad, CA 920

1

0-

7226

USA

Tel 760-438-2021
Fax 800-869-7668 / 760-931-9189
www.myronl.com

pH Conductivity Instrumentation

Accuracy • Reliability • Simplicity

© Myron L Company 2001

(W)

Printed in U.S.A.

Specifications subject to change without notice

This manual is related to the following products: