beautypg.com

Description – Myron L M6/PH User Manual

Page 2

background image

DESCRIPTION

This manual describes your Myron L pDS meter, tells you how to

use it, and how to keep it working accurately for many years.

Myron L pDS Meters are compact, light-weight, self-contained

conductivity/pH meters. They contain a built-in conductivity cell with
a replaceable pH sensor. Models EP11/pH and M6/pH have dials
calibrated in micromhos (equivalent to the metric microsiemens) and
in pH units. They will quickly determine the conductivity and pH of
almost any solution.

Both are 3.4" x 4.5" x 4.0" (85 x 129 x 126mm) and weigh less

than one pound (0.45 kg). pDS Meters are completely self-contained.
The built-in conductivity cell is automatically temperature
compensated from 50

°

to 160

°

F (10

°

to 71

°

C). They are powered by

two 9 volt transistor batteries good for at least 2000 tests, or one
year shelf life.

MODEL

CONDUCTIVITY RANGES

pH RANGE

EP11/pH

0-10, 0-100, 0-1000,
0-10,000 micromhos (

µ

mhos)

2-12 pH

M6/pH

0-5000 micromhos (

µ

mhos)

2-12 pH

ACCURACY: (Both Models)

Conductivity:

±

2% of full scale

pH:

±

0.2 pH units

REPEATABILITY: (Both Models)

±

1%

By using an RE-10 Range Extender (see

ACCESSORIES), the

maximum conductivity range of each meter may be increased ten
times.

The pictures in this manual show the major operating parts of

your Myron L DS Meter. Handle your DS Meter and identify these
parts to become familiar with it.

IMPORTANT: The pH sensor in your pDS Meter has a limited life
much like a battery. Sensor longevity is affected both by the types of
solution tested and care given your meter. pH sensors are user
replaceable. See MAINTENANCE for additional information.