Hot key 2. wrt protect menu 2. enable wrt 10min, Open “ wrt protect menu ” in the hot key menu, 6 7 8 9 a – APC AXFA11G User Manual
Page 93

IM 01E20C01-01E
8-19
8. OPERATION VIA HART COMMUNICATOR
Changing Password
Example: Change the password from “1 2 3 4” to “6 7 8 9 A”.
F0822-1.EPS
ABORT
ABORT
OK
AXFA11 : FI-1201
Enter current
password to enable to
write for 10
minutes:
Hot key
2. Wrt Protect Menu
2. Enable Wrt 10min
HELP
DEL
ABORT
ENTER
1
2
AXFA11 : FI-1201
If you wish to
release completely,
you have to change
password to all of
spaces.
HELP
DEL
ABORT
OK
3
Open “Wrt protect menu” in the Hot key Menu.
AXFA11 : FI-1201
Release the write
protection for 10
minutes.
DEL
DEL
ABORT
OK
AXFA11 : FI-1201
Wrt Protect Menu
1 Write protect No
2 Enable Wrt 10min
3 New Password
4 Software seal Keep
HELP
SAVE
ABORT
OK
4
5
AXFA11 : FI-1201
Enter new password to
change state of write
protect
HELP
DEL
ABORT
ENTER
F4
(OK)
F4
(OK)
F4
(ENTER)
YZ/
3
6
AXFA11 : FI-1201
Re-enter new Password
write 30 seconds:
6789A
6789A
HELP
DEL
ABORT
ENTER
7
AXFA11 : FI-1201
Change to New
password
HELP
DEL
F4
8
(OK)
AXFA11 : FI-1201
It changed the state
of protection related
password.
HELP
DEL
OK
F4
(ENTER)
“1 2 3 4”
Enter the password and press ENTER (F4).
Press OK (F4).
Write protect status is released for 10 minutes.
Press OK (F4).
Select “New Password”.
Enter “6 7 8 9 A” and press ENTER (F4).
Enter “6 7 8 9 A” again and press ENTER (F4).
Press OK (F4).
“6 7 8 9 A”
F4
(ENTER)
“6 7 8 9 A”
F0822-2.EPS
or