LI200X-LC PYRANOMETER FOR METDATA1

3

TABLE 4-1. Multipliers Required for Average

Flux and Total Flux Density in Sl and

English Units

UNITS

MULTIPLIER

PROCESS

W m

-2

200

Average

MJ m

-2

t * 0.0002

Total

kJ m

-2

t * 0.2

Total

cal cm

-2

min

-1

0.2 * (1.4333)

Average

cal cm

-2

t * 0.2 * (0.02389)

Total

t = datalogger execution interval in seconds

EXAMPLE 1. Sample Instructions used to Measure an Average Flux

;{CR10X}
;
*Table 1 Program

01:

10

Execution Interval (seconds)

01:

Volt (Diff) (P2)
1:

1

Reps

2:

22

± 7.5 mV 60 Hz Rejection Range

3:

3

DIFF Channel

4:

1*

Loc [ W_m2 ]

5:

200***

Mult

6:

0

Offset

;Set negative values to zero.
;
02:

If (X<=>F) (P89)
1:

1*

X Loc [ W_m2 ]

2:

4

<

3:

0

F

4:

30

Then Do

03:

Z=F (P30)
1:

0

F

2:

0

Exponent of 10

3:

1*

Z Loc [ W_m2 ]

04:

End (P95)

05:

If time is (P92)
1:

0

Minutes (Seconds --) into a

2:

30

Interval (same units as above)

3:

10

Set Output Flag High (Flag 0)

06:

Real Time (P77)
1:

0110

Day,Hour/Minute