beautypg.com

2 cr10x example program – Campbell Scientific LI190SB LI-COR Quantum Sensor User Manual

Page 13

background image

LI190SB Quantum Sensor

5.2.2 CR10X Example Program

;{CR10X}
*Table 1 Program
01: 10.0000 Execution

Interval

(seconds)


;Measure LI190SB
1: Volt (Diff) (P2)
1: 1

Reps

2: 23

25 mV 60 Hz Rejection Range ;15 mV for 21X, CR7; 10 mV for CR23X

3: 1

DIFF

Channel

4: 3

Loc [ PAR_Den ]

5: 1

Multiplier

6: 0

Offset


;Set negative values to zero
2: If (X<=>F) (P89)
1: 3

X Loc [ PAR_Den ]

2: 4

<

3: 0

F

4: 30

Then

Do


3: Z=F x 10^n (P30)
1: 0

F

2: 0

n, Exponent of 10

3: 3

Z Loc [ PAR_Den ]


4: End (P95)

;Convert mV to mmoles m

-2

for 10 second scan

5: Z=X*F (P37)
1: 3

X Loc [ PAR_Den ]

2: 2.56686 F

;Will differ for each sensor and scan rate

3: 4

Z Loc [ PAR_Tot ]


;Convert mV to µmoles s

-1

m

-2

6: Z=X*F (P37)
1: 3

X Loc [ PAR_Den ]

2: 256.686 F

;Will differ for each sensor

3: 3

Z Loc [ PAR_Den ]


7: If time is (P92)
1: 0

Minutes (Seconds --) into a

2: 60

Interval (same units as above)

3: 10

Set Output Flag High (Flag 0)


8: Set Active Storage Area (P80)
1: 1

Final Storage Area 1

2: 101

Array

ID


9: Real Time (P77)
1: 1220

Year,Day,Hour/Minute (midnight = 2400)

9