Campbell Scientific IRTS-P Precision Infrared Temperature Sensor User Manual
Page 10

IRTS-P Precision Infrared Temperature Sensor
6
;Calculate correction factor (SEC)
7: Z=1/X (P42) ; {1/Psb}
1: 4
X Loc [ Psb ]
2: 4
Z Loc [ Psb ]
8: Z=X*F (P37) ; {.25/Psb}
1: 4
X Loc [ Psb ]
2: 0.25
F
3: 4
Z Loc [ Psb ]
9: Z=X-Y (P35) ; {ATT - Hsb}
1: 2
X Loc [ AppTargT ]
2: 5
Y Loc [ Hsb ]
3: 5
Z Loc [ Hsb ]
10: Z=X*Y (P36) ; {ATT - Hsb}^2
1: 5
X Loc [ Hsb ]
2: 5
Y Loc [ Hsb ]
3: 5
Z Loc [ Hsb ]
11: Z=X-Y (P35) ; {subtract Ksb}
1: 5
X Loc [ Hsb ]
2: 6
Y Loc [ Ksb ]
3: 6
Z Loc [ Ksb ]
12: Z=X*Y (P36) ; {calculate SEC}
1: 4
X Loc [ Psb ]
2: 6
Y Loc [ Ksb ]
3: 7
Z Loc [ SEC ]
;Calculate corrected target temperature (CTT)
13: Z=X-Y (P35) ;
1: 2
X Loc [ AppTargT ]
2: 7
Y Loc [ SEC ]
3: 8
Z Loc [ CTT ]
14: If time is (P92)
1: 0
Minutes (Seconds --) into a
2: 60
Interval (same units as above)
3: 10
Set Output Flag High (Flag 0)
15: Real Time (P77)
1: 1220
Year,Day,Hour/Minute
(midnight
=
2400)
16: Average (P71)
1: 3
Reps
2: 1
Loc [ RefTemp ]
17: Average (P71)
1: 1
Reps
2: 8
Loc [ CTT ]