Campbell Scientific HMP60 Temperature and Relative Humidity Probe User Manual
Page 22

HMP60 Temperature and Relative Humidity Probe
7.6.2 Sample CR10(X) Program that Computes Vapor Pressure
and Saturation Vapor Pressure
;Measure the HMP60 temperature.
;
01: Volt (SE) (P1)
1: 1
Reps
2: 5
2500 mV Slow Range
;CR510 (2500 mV); CR23X (1000 mV); 21X,
CR7
(5000
mV)
3: 3
SE
Channel
;Black wire (SE 3), Blue wire (G)
4: 1
Loc [ T_C ]
5: .1
Mult
;See Table 7-2 for alternate multipliers
6: -40
Offset
;See Table 7-2 for alternate offsets
;Measure the HMP60 relative humidity.
;
02: Volt (SE) (P1)
1: 1
Reps
2: 5
2500 mV Slow Range
;CR510 (2500 mV); CR23X (1000 mV); 21X,
CR7
(5000
mV)
3: 4
SE
Channel
;White wire (SE 4), Blue wire (G)
4: 2
Loc [ RH_frac ]
5: .001
Mult
;See Table 7-3 for alternate multipliers
6: 0
Offset
;Limit the maximum value of relative humidity
;to 1 (expressed as a fraction).
;
03: If (X<=>F) (P89)
1: 2
X Loc [ RH_frac ]
2: 3
>=
3: 1
F
4: 30
Then
Do
04: Z=F (P30)
1: 1
F
2: 0
Exponent
of
10
3: 2
Z Loc [ RH_frac ]
05: End (P95)
;Compute the saturation vapor pressure in kPa.
;The temperature must be in degrees Celsius.
;
06: Saturation Vapor Pressure (P56)
1: 1
Temperature Loc [ T_C ]
2: 3
Loc [ e_sat ]
;Compute the vapor pressure in kPa.
;Relative humidity must be a fraction.
;
07: Z=X*Y (P36)
1: 3
X Loc [ e_sat ]
2: 2
Y Loc [ RH_frac ]
3: 4
Z Loc [ e ]
16