beautypg.com

Campbell Scientific HMP60 Temperature and Relative Humidity Probe User Manual

Page 22

background image

HMP60 Temperature and Relative Humidity Probe

7.6.2 Sample CR10(X) Program that Computes Vapor Pressure

and Saturation Vapor Pressure

;Measure the HMP60 temperature.
;
01: Volt (SE) (P1)
1: 1

Reps

2: 5

2500 mV Slow Range

;CR510 (2500 mV); CR23X (1000 mV); 21X,

CR7

(5000

mV)

3: 3

SE

Channel

;Black wire (SE 3), Blue wire (G)

4: 1

Loc [ T_C ]

5: .1

Mult

;See Table 7-2 for alternate multipliers

6: -40

Offset

;See Table 7-2 for alternate offsets


;Measure the HMP60 relative humidity.
;
02: Volt (SE) (P1)
1: 1

Reps

2: 5

2500 mV Slow Range

;CR510 (2500 mV); CR23X (1000 mV); 21X,

CR7

(5000

mV)

3: 4

SE

Channel

;White wire (SE 4), Blue wire (G)

4: 2

Loc [ RH_frac ]

5: .001

Mult

;See Table 7-3 for alternate multipliers

6: 0

Offset


;Limit the maximum value of relative humidity
;to 1 (expressed as a fraction).
;
03: If (X<=>F) (P89)
1: 2

X Loc [ RH_frac ]

2: 3

>=

3: 1

F

4: 30

Then

Do


04: Z=F (P30)
1: 1

F

2: 0

Exponent

of

10

3: 2

Z Loc [ RH_frac ]


05: End (P95)

;Compute the saturation vapor pressure in kPa.
;The temperature must be in degrees Celsius.
;
06: Saturation Vapor Pressure (P56)
1: 1

Temperature Loc [ T_C ]

2: 3

Loc [ e_sat ]


;Compute the vapor pressure in kPa.
;Relative humidity must be a fraction.
;
07: Z=X*Y (P36)
1: 3

X Loc [ e_sat ]

2: 2

Y Loc [ RH_frac ]

3: 4

Z Loc [ e ]

16