beautypg.com

Installation, 1 siting, 2 mounting and assembly – Campbell Scientific HMP60 Temperature and Relative Humidity Probe User Manual

Page 16

background image

HMP60 Temperature and Relative Humidity Probe

7. Installation

7.1 Siting

Sensors should be located over an open level area at least 9 m (EPA) in
diameter. The surface should be covered by short grass, or where grass does
not grow, the natural earth surface. Sensors should be located at a distance of
at least four times the height of any nearby obstruction, and at least 30 m
(EPA) from large paved areas. Sensors should be protected from thermal
radiation, and adequately ventilated.

Standard measurement heights:

1.5 m (AASC)
1.25 – 2.0 m (WMO)
2.0 m (EPA)

See Section 10 for a list of references that discuss temperature and relative
humidity sensors.

7.2 Mounting and Assembly

Pull off the yellow shipping cap (see Figure 4-1 in Section 4, Quickstart).

The HMP60 must be housed inside a solar radiation shield when used in the
field. The HMP60 is typically housed in the 41303-5A 6-Plate Radiation
Shield (Figures 4-2 and 4-3 in Section 4, Quickstart). The HMP60 is held
within the 41303-5A by a mounting clamp (Figure 4-4 in Section 4,
Quickstart).

This probe may also be housed in a 41003-5 10-plate shield if additional
hardware is used. A 41322 Adapter Plate allows the HMP60 to be mounted in
the lower part of the 41003-5 10-plate shield. Alternatively, a 41381 extension
tube and the 6637 split nut plug can be used to mount the HMP60 in a higher
part of the 41003-5; the #41381 cable is also required.

Both the 41303-5A and the 41003-5 attach to a crossarm, mast, or user-
supplied pipe with a 1.0-in. to 2.1-in. outer diameter.

10