CS100 Barometric Pressure Sensor

TABLE 7-1. Signal and Ground Connectors for CS100

Wire

CS100 Terminal

Datalogger

Single-Ended Measurement

Datalogger

Differential Measurement

Blue

VOUT

S.E. Input

High Side of Diff Input

Yellow AGND

AG (CR10(X), CR500, CR510)

(Other Dataloggers)

Low Side of Diff. Input

Black GND

(21X, CR7, CR9000(X))

G (Other Dataloggers)

(21X, CR7, CR9000(X))

G (Other Dataloggers)

Green EXT

TRIG

Control port (use to turn power
on/off)

Control port (use to turn power
on/off)

Red

SUPPLY

12 VDC

12 VDC

Shield Shield

G (CR10(X), CR500, CR510)

(Other Dataloggers)

G (CR10(X), CR500, CR510)

(Other Dataloggers)

Improper wiring may damage the CS100 beyond repair.

WARNING

7.2.2 5-pin Screw Terminal Plug Connector

The datalogger connects to the CS100 via a 5-pin screw terminal plug
connector. This connector is removable and may be replaced. The
replacement connector may come with a connector key attached to it to ensure
that the connector is plugged into the CS100 right side up (see FIGURE 7-3).
When the connector is right side up, it will easily plug into the barometer.

FIGURE 7-3. Connector key attached to 5-pin screw terminal plug

connector

A 5-pin screw terminal that is plugged in upside down
will damage the sensor—perhaps beyond repair.

WARNING