CR10XTCR Thermocouple Reference

6

TABLE 4. Polynomial Coefficients

Coefficient

Value

C0

-53.4601

C1

90.807

C2

-83.257

C3

52.283

C4

-16.723

C5

2.211

6. Electrically Noisy Environments

AC power lines can be the source of electrical noise. If the datalogger is in an
electronically noisy environment, the CR10XTCR temperature measurement
should be measured with the AC half bridge (Instruction 5) with the 60 Hz
rejection integration option on the CR10X (see Section 13 of the datalogger
manual for more information on noise). Instruction 11's fast integration will not
reject 60 Hz noise.

Example 2. Sample CR10X Instructions Using AC Half Bridge

01: AC Half Bridge (P5)

01: 1

Rep

02: 22

7.5 mV 60 Hz rejection Range

03: 1

IN Chan

04: 3

Excite all reps w/EXchan 3

05: 2000

mV Excitation

06: 15*

Loc [:Ref_Temp]

07: 800

Mult

08: 0

Offset

02: Polynomial (P55)

01: 1

Rep

02: 15*

X Loc Air Temp

03: 15*

F(X) Loc [:Ref_Temp]

04: -53.46

C0

05: 90.807

C1

06: -83.257

C2

07: 52.283

C3

08: -16.723

C4

09: 2.211

C5

* Proper entries will vary with program and input location assignments.