beautypg.com

Installation and wiring, Programming, Instruction 11 details – Campbell Scientific CR10TCR Thermocouple Reference User Manual

Page 6

background image

CR10TCR THERMOCOUPLE REFERENCE

2

3. INSTALLATION AND WIRING

The wiring diagram for the CR10TCR is shown
in Figure 1. The probe requires a single-ended
analog input channel and an excitation channel.
After the CR10TCR and thermocouple are
connected to the datalogger, install the wiring
panel cover (Figure 2 and 3).

FIGURE 2. Wiring Diagram

FIGURE 3. Wiring Panel, CR10TCR, and

Wiring Panel Cover

FIGURE 4. CR10TCR Installed on the CR10

Wiring Panel

4. PROGRAMMING

Instruction 11 is used to measure the CR10TCR.
Instruction 11 provides AC excitation, makes a
single ended voltage measurement, and
calculates temperature with a fifth order
polynomial. A multiplier of 1.0 and an offset of

0.0 yields temperature in Celsius. For Fahrenheit,
use a multiplier of 1.8 and an offset of 32.

CAUTION: The reference temperature for
thermocouples must be in degrees Celsius.

EXAMPLE 1. Sample CR10 Instructions

01:

P11

Temp 107 Probe

01:

1

Rep

02:

1

IN Chan

03:

3

Excite all reps w/EXchan 3

04:

15*

Loc [:Ref_Temp]

05:

1

Mult

06:

0

Offset

*

Proper entries will vary with program and

input location.

5. INSTRUCTION 11 DETAILS

Understanding the details in this section are not
necessary for general operation of the CR10TCR
with Campbell Scientific's dataloggers.

Instruction 11 outputs a precise 2 VAC excitation and
measures the voltage drop due to the sensor
resistance (Figure 5). The thermistor resistance
changes with temperature. Instruction 11 calculates
the ratio of voltage measured to excitation voltage
(Vs/Vx) which is related to resistance, as shown below:

Vs/Vx = 1000/(Rs+249000+1000)

where Rs is the resistance of the thermistor.

See the measurement section of the datalogger
manual for more information on bridge
measurements.

Instruction 11 then calculates temperature using a
fifth order polynomial equation correlating Vs/Vx with
temperature. The polynomial coefficients are given
in Table 3. The polynomial input is (Vs/Vx)*800.
Resistance and datalogger output at several
temperatures are shown in Table 4.

Black

Red

Clear

FIGURE 5. CR10TCR Reference Temperature

Schematics

Analog
Ground; Clear

Switched
Excitation;
Black (E3)

Single-Ended Input; Red (1H)