CM3 Pyranometer

5

If a 21X is used to measure the CM3 and it powers a 12 VDC sensor or 12
VDC radio, the current drawn off the 12 VDC supply may cause a difference in
ground potential between the 21X ground terminals and the reference ground
point in the datalogger. This ground potential results in an offset on single
ended measurements. This offset can be as large as +60 mV. Thus, single
ended measurements should be avoided. The offset does not, however, affect
differential measurements. While the 21X is the logger most susceptible to this
condition, the other dataloggers can be affected if the amount of power is too
large or the power return line is connected to analog ground (AG).

FIGURE 4-1. CM3 Wiring

Color

Function

CR10(X),
CR510

21X/CR7

CR23X

Red

Signal

Differential
Channel - H

Differential
Channel - H

Differential
Channel - H

Blue

Signal Reference

Differential
Channel - L

Differential
Channel - L

Differential
Channel - L

White

Signal Ground

AG

Clear

Shield

G

White = Signal Ground

Blue = Signal Reference

Red =

Signal

Clear = Shield