3WHB10K 3-Wire Half Bridge Terminal Input Module

2

3. Wiring

Shield

A

B

H

L

G

10 k

Datalogger

Vx

H

L

or AG

or G

FIGURE 3-1. 3-Wire Half Bridge Used to Measure PRT

TABLE 3-1. 3WHB10K Connections to

Campbell Scientific Dataloggers



Function



Label/Lead


CR10X,
CR510

CR23X,
CR1000,
CR800, CR850,
CR3000


21X, CR7,
CR9000X

Excitation

Black Wire

E1

EX1

Excitation 1

V1 Reference

H

SE1

SE1

1H

V2 Sense

L

SE2

SE2

1L

Ground G

AG

4. Programming Examples

The following examples simply show the two instructions necessary to 1)
make the measurement and 2) calculate the temperature. The result of the 3-
wire half bridge measurement as shown is R

s

/R

o

, the input required for the

PRT algorithm to calculate temperature.

All the examples are for a 100 Ohm PRT in the 3WHB10K. The excitation
voltages used were chosen with the assumption that the temperature would not
exceed 50°C. Table 4-1 lists excitation voltage as a function of maximum
temperature and the input voltage ranges used with the different dataloggers.
Calculation of optimum excitation voltage is discussed in Section 5.1.

The multiplier shown is for a 100 Ohm PRT. The multiplier for a 1000 Ohm
PRT is 10.