Aalborg XFM Digital User Manual
Page 34

31
COMMAND
NAME
DESCRIPTION
No.
COMMAND SYNT
A
X
Command
Argument 1
Argument 2
Argument 3
Argument 4
Response
Auto
Zero
Starts /reads the status of
the auto zero feature (Note:
The Z,N command can be
used only when absolutely
no flow thru the meter and
n
o
e
ar
lie
r
then 6 minutes after
power
up. It can take
several
minutes to complete. U
n
it
w
ill
not respond to other commands
when this is in progress.)
5
Z
N
(do it now)
ZN
W
(Write Zero to
EEPROM)
ZW (when done)
S
(status while auto
zero in progress)
ZNI,
is in progress.
V
(Display zero value)
ZV
,
Flow
Alarms
Sets / reads the status of
the gas flow alarms.
Note:
High and Low limits
have to be entered in the
%F
.S. High alarm value
has to be more than Low
alarm value.
Alarm conditions:
Flow > High Limit = H
Flow < Low Limit = L
Low < Flow < High = N
6
A
H
(high flow limit)
alue> (0-100%FS) AH alue> L (low flow limit) alue> (0-100%FS) AL alue> A (action delay in seconds) alue> (0-3600 sec.) AA: alue> E (enable alarm) AE D (disable alarm)* AD R (read current status) N (no alarm) H (high alarm) L (low alarm) S (Read current settings) AS:M,L,H,D,B where: M – mode (E/D) L – Low settings (%FS) H – High settings (%FS) D – Action Delay (sec) B – Latch mode (0-3) B Block (Latch) mode alue> (0-disabled*) (1-enabl’d L) (2-enabl’d H) (3 –both L,H) AB: alue> where: V alue = 0 - 3