2. output example – A&D Ej Series Compact Balance EJ-2000 User Manual
Page 33
31
10-2. Output example
To print the GLP report, set the function “info 1” and use MODE 3 of the printer
AD-8121B.
To output the GLP report to a personal computer, set the function “info 2”.
Data format for “calibration report”
Perform calibration using a weight. Then the balance will output a calibration report.
See “7-1. Calibration using a weight” about the calibration.
AD-8121 format “info 1”
General format “info 2”
Data format for “calibration test report”
Perform calibration test using a weight. Then the balance will output a calibration test
report. See “10-3. Calibration test using a weight
” about the calibration test.
AD-8121 format “info 1”
General format “info 2”
:Space,
ASCII
20h
CR :Carriage return, ASCII 0Dh
LF :Line feed, ASCII 04Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration
executed
Calibration weight
Signature
A & D
MODEL EJ-300
S/N 1234567
ID ABCDEF
DATE 09/03/21
02:53:21 PM
CALIBRATED(EXT.)
CAL.WEIGHT
+300.00 g
SIGNATURE
- - - - - - - -
~~~~~~~~~~~
A
~
&
~
D
MODEL
~~~~~
EJ-300
S/N
~~~~~~
1234567
ID
~~~~~~~~
ABCDEF
DATE
TIME
CALIBRATED(EXT.)
CAL.WEIGHT
~~~~~~
+300.00
~~
g
SIGNATURE
-
~
-
~
-
~
-
~
-
~
-
~
-
~
-
~
A & D
MODEL EJ-300
S/N 1234567
ID ABCDEF
DATE 09/03/21
03:15:40 PM
CALIBRATED(EXT.)
ACTUAL
0.00 g
+299.99 g
TARGET
+300.00 g
SIGNATURE
- - - - - - - -
~~~~~~~~~~~
A
~
&
~
D
MODEL
~~~~~
EJ-300
S/N
~~~~~~
1234567
ID
~~~~~~~~
ABCDEF
DATE
TIME
CAL.TEST(EXT.)
ACTUAL
~~~~~~~~~
0.00
~~
g
~~~~~~
+229.99
~~
g
TARGET
~~~~~~
+300.00
~~
g
SIGNATURE
-
~
-
~
-
~
-
~
-
~
-
~
-
~
-
~
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration
test
Signature
Zero value
Weight value
Target weight value