beautypg.com

Func unit unit, Unit end id – A&D Ej Series Compact Balance EJ-2000 User Manual

Page 21

background image

19

8-4. Storing weighing units

It is possible to store the weighing units that will be actually used from the units
available. For the units available, see “5-3. Units”.
Select and store the weighing units as described below:

1. Press and hold the SAMPLE key to display

func .





2. Press the SAMPLE key several times to display

Unit .





3. Press the PRINT key.




4. Press the SAMPLE key to display a weighing unit

to be stored.





5. Press the RE-ZERO key to select the weighing unit.

The selected weighing unit is shown with the STABLE
indicator.

‰

At this stage, the weighing unit is not stored in memory.



6. Repeat steps 4. and 5. to store all weighing units to be

used.


7. Press the PRINT key to store the selected weighing

units in memory.

id appears after end .


8. Press the UNITS key to return to the weighing

mode.

‰

Note

When the balance is switched on, it starts with
the unit that was stored first at step 5.



func

Unit

Unit

Press and hold

Unit

end

id

Returns to

the weighing mode

Each pressing switches

the units available in the

order described on 5-3.

To select

Press

several times

To store