Chapter 4 storage and transportation, Storage, Chapter 4 – Vaisala RS92-SGP User Manual
Page 45: Storage and transportation
Chapter 4 __________________________________________________ Storage and Transportation
VAISALA_______________________________________________________________________ 43
CHAPTER 4
STORAGE AND TRANSPORTATION
This chapter provides information for the transport and storage of the
product.
Storage
Radiosondes must be stored and used properly in accordance with
applicable instructions, the User’s Guide, and specifications issued by
Vaisala.
Proper storage conditions must fulfill the following requirements:
Radiosondes must be kept in their original packaging (unopened
vacuum envelopes) in a dry, ventilated indoor storage space, and within
the following key environmental limits (ref. IEC 60721-3-1 class 1K2):
-
Temperature +5 °C to +40 °C
-
Relative humidity below 85%
Vaisala AUTOSONDE storage compartment requirements are:
-
The maximum number of days a radiosonde can be loaded in
AUTOSONDE is 24
-
Relative humidity below 50 %
-
Temperature +15 °C to +35 °C