Chapter 3_____________________________________________________________________ Operation
VAISALA _____________________________________________________________________________7
6. Let the calibrator settle for 10 minutes.
7. Connect the reference gas and adjust the flow rate to 0.6 l/min
with a screwdriver (item 4, Figure 1). Flush the chamber with the gas
for 5 minutes.
8. Input the ambient barometric pressure.
MF_PRESSURE
XXXX
<
CR
>
where xxxx= pressure in hPa.
Example:
>MF_PRESSURE 1013
MF_PRESSURE=1013
>
9. Input the ambient temperature.
MF_TEMP
XXX
<
CR
>
where xxx = temperature is given in tenths of °C (25 °C = 250).
Example:
>MF_TEMP 250
MF_TEMP=250
>
10. List the carbon dioxide concentration.
MF_MODE 2<
CR
>
Example:
>MF_MODE 2
MF_MODE=2
1620.1
1618.3
1619.8
1619.8
...
The carbon dioxide concentration (in ppm) is measured every 0.5
seconds.
- DM500 (138 pages)
- DM70 (93 pages)
- DMT132 (74 pages)
- DMT143 (76 pages)
- DMT152 (70 pages)
- DMT242 (4 pages)
- DMT340 (191 pages)
- DMT345 (185 pages)
- DPT145 (63 pages)
- DPT146 (71 pages)
- PTU300 (217 pages)
- PTB330TS (89 pages)
- PTB220 (10 pages)
- PTB220 (113 pages)
- PTB330 (144 pages)
- PTU200 (64 pages)
- PTU200MIK1 (18 pages)
- SPH10 (2 pages)
- SPH20 (2 pages)
- PTB110 (4 pages)
- PTB200 (30 pages)
- PTB210 (analog) (27 pages)
- PTB210 (serial) (32 pages)
- GM70 (68 pages)
- GMD20 (4 pages)
- GML20 (2 pages)
- GML20T (2 pages)
- GMM20W (5 pages)
- GMM220 (6 pages)
- GMP231 (90 pages)
- GMP231 (2 pages)
- GMP343 (94 pages)
- GMR20 (2 pages)
- GMT220 (42 pages)
- GMW90 (101 pages)
- XMW90 (4 pages)
- MM70 (67 pages)
- MM70 (71 pages)
- MMT162 (66 pages)
- MMT310 (81 pages)
- MMT330 (181 pages)
- MMT330 (171 pages)