Chapter 3_____________________________________________________________________ Operation

VAISALA _____________________________________________________________________________7

6. Let the calibrator settle for 10 minutes.

7. Connect the reference gas and adjust the flow rate to 0.6 l/min
with a screwdriver (item 4, Figure 1). Flush the chamber with the gas
for 5 minutes.

8. Input the ambient barometric pressure.

MF_PRESSURE

XXXX

<

CR

>

where xxxx= pressure in hPa.

Example:

>MF_PRESSURE 1013
MF_PRESSURE=1013
>

9. Input the ambient temperature.

MF_TEMP

XXX

<

CR

>

where xxx = temperature is given in tenths of °C (25 °C = 250).

Example:

>MF_TEMP 250
MF_TEMP=250

>

10. List the carbon dioxide concentration.

MF_MODE 2<

CR

>

Example:

>MF_MODE 2
MF_MODE=2
1620.1
1618.3
1619.8
1619.8
...

The carbon dioxide concentration (in ppm) is measured every 0.5
seconds.