New calibration parameters are now saved in the probe.

14. Disconnect the gas flow and remove the probe from the chamber.

After two-point calibration it is recommended to ensure the correct operation of
the probe by carrying out a spot checking procedure using both low and high end
gases. The readings should be within the accuracy specifications of the
transmitter.