beautypg.com

Technical data, Technical specifications, Operating range – Vaisala PTB210 (analog) User Manual

Page 21: Accuracy, Technical specifications operating range, Chapter 5

background image

Chapter 5 _____________________________________________________________ Technical Data

VAISALA________________________________________________________________________ 19

CHAPTER 5

TECHNICAL DATA

Technical Specifications

Operating Range

Pressure range (order specified)

500 ... 1100 hPa

800 ... 1060 hPa

600 ... 1060 hPa

900 ... 1100 hPa

Operating temperature range

-40 ... +60

°C

Humidity range

non condensing

Accuracy

Non linearity*

± 0.20 hPa

Hysteresis*

± 0.05 hPa

Repeatability*

± 0.05 hPa

Calibration uncertainty**

± 0.15 hPa

Total accuracy (20

°C)***

± 0.30 hPa


Temperature dependence
(-40 ... +60

°C)****

± 0.50 hPa

Total accuracy***
(-40 ... +60

°C) ± 0.60 hPa