ASSEMBL

ING THE CASE

AND AN

EXISTING BAROMETER

_________________________________________

V

AISALA

______________________________________________________________________________5

ASSEMBLING THE CASE AND AN
EXISTING BAROMETER

To assemble the existing PTB220 digital barometer and the wooden
case follow the instructions below.

NOTE! It is recommended to remove the cover hinge of the barometer
by loosening its screws. Then the cover can easily be detached after the
barometer is installed into the case.

1. Remove the wooden tilting frame from

the case.

2. Unscrew the base plate.
3. Attach the PTB220 onto the base with

four M6x16 screws.

4. Thread the flat cable between the

battery covering and the base plate.
Connect the cable to D-terminal.

5. Remove the original 1/8" pressure

fitting of the barometer. Then, attach

the pressure tubing connector to the

PTB220. It is recommended to use
Loctite 572 joint paste for sealing.

6. Place the assembly into the tilting

frame. Check that the flat cable and
the pressure tubing come out of the

back opening.

7. Attach the base with six screws.