beautypg.com

Sper Scientific 800043 Ultimate Thermometer User Manual

Page 6

background image

6

G. LCD BACK LIGHT - Press the BACK LIGHT >2S5 button (5) for at

least 2 seconds. The LCD will light for approximately 10 seconds, then
turn off.

H. AUTO POWER OFF - If no buttons are pushed for approximately

10 minutes, the meter will automatically turn off. To disable this feature,
select the Data Record function (page 5) by pressing the REC button
(6). "REC" will be displayed.

I. BATTERY REPLACEMENT - Accurate thermometer

readings may be taken for several hours after the low battery
indicator is displayed. Open the battery cover, install a fresh
9V battery and replace the cover.

J. RS232 PC SERIAL INTERFACE - This unit features an RS232
Output Socket
(9) on the side of the unit. Gently lift the hinged cover

from the bottom. The output is a 16 digit data stream which can be
utilized by user's specific application. An RS232 lead with the following
connection is required to link the instrument with the PC serial input:







RS232 FORMAT: 9600, N, 8, 1

The 16 digit data stream will be displayed as follows:

D15D14D13D12D11D10D9D8D7D6D5D4D3D2D1D0

D0

End Word

D1 & D8

Display reading, DI " LSD, D8 MSD. For example: If the
display reading is 1234, then D8 to D I is: 00001234

D9

Decimal Point(DP), position from right to the left
0 = No DP, 1 = 1 DP, 2 = 2 DP, 3 = 3 DP

D10

0 = Positive, 1 = Negative

D11 & D12

Anunuciator for Display: 01=°C, 02=°F

D13

1

D14

4

D15

Start Word

Each digit indicates the following status:

Meter (3.5 mm jack plug) PC (9W "D" Connector)

Center Pin Pin 4 Pin 2 2.2K
Ground/Shield Pin 2 Pin 5 resister