Sper Scientific 840018 Sound Level Meter Pen User Manual

Page 5

background image

- 5 -

electrical calibration of the unit.

• The LCD displays "CAL" for about 1

second. As needed, use the calibration
tool to adjust the Calibration (CAL) VR
until "94.0dB" is displayed.

• This ends the calibration. Make sure

that the Measurement Slide Switch is
returned to the "MEAS" (up) position
and the Battery Cover is reattached.


2. Acoustic calibration using a standard

sound calibrator such as Sper Scien-
tific's 840031.

• With the Measurement Slide Switch in

the "MEAS" (up) position, place the mi-
crophone in the calibrator.

• As needed, use the calibration tool to

adjust the Calibration (CAL) VR until
"94.0dB" is displayed.


Measurement
1. Press the Power / Max Hold button to

turn the meter on.

2. Point the Microphone at the noise

source.

3. The LCD displays the measured sound

level. Readings are updated twice
every second.
• "OVER" is displayed when the sound