Sper Scientific 840018 Sound Level Meter Pen User Manual

Page 4

background image

- 4 -

III. Operating Instructions

Precautions and Maintenance
• Do not store or operate the instrument at

high temperatures or humidity.

• Keep the microphone dry and avoid se-

vere shock or vibration.

• Wind blowing across the microphone can

cause measurement errors, these effects
are reduced by using the windscreen.

• Remove the batteries if the unit is not in

use for a long period of time.

• Wipe the case with a dry cloth, do not

use abrasives or solvents.

Calibration

Calibrate the meter before first use or
when the reading appears inaccurate.
If you require a certificate of calibration
traceable to NIST, the unit may be sent to
Sper Scientific for servicing (fee applies).

There are two methods of calibration:
1. Internal Electrical Calibration
• With the unit on, lightly press the

Battery Cover to slide it down and off.

• Move the Measurement Slide Switch

down. This activates the built-in
oscillator (94dB 1 kHz, sine wave) for