8

SPECIFICATIONS

Display:

3.5 digits LCD, maximum reading
1999

Range:

200/2000 mG,20/200µT.

Resolution:

0.1/1 mG or 0.01/0.1 µT.

Frequency
response:

30Hz to 300Hz.

Sensor:

Single Axis

Accuracy:

±(2.5% + 6 dgt) at 50Hz/60Hz.

Over load:

LCD display “OL”.

Sample rate:

2.5 times per second.

Battery:

9V Battery

Battery life:

Approximate 100 hours.

Operating
temperature &
humidity:

5ºC to 40ºC, below 80% RH.

Storage
temperature &
humidity:

-10º C to 60ºC, below 70%. RH

Weight:

170g.

Dimensions:

130x56x38 mm.

Accessories:

User’s manual, 9V battery, Carrying
case

5

PANEL DESCRIPTION

1. Sensor .
2. LCD
3. Power on/off.
4. Maximum hold button
5. Data hold button
6. mG unit select button
7. µT unit select button
8. Range select button
9. Battery compartment

MEASUREMENT PROCEDURES

1. Press to turn on the unit.

2. Select for mG or for µT.

3. Position the front (see Figure 1) of the meter to measure

the electromagnetic waves. Change the measurement
angle or position to obtain the highest reading value
(see Figure 2 through Figure 6).

4. Read the measured value. The display of “OL” indicates

an overload. Press to select a higher range and
repeat the measurement.