The meter can record up to 16,000 data points, stored in a maximum of
250 memory blocks. All data points saved during one data logging ses-
sion are stored in one memory block.

The 16 digits data stream will be displayed in the following format:

D0

End Word = 0D

D1 & D8

Display reading, Dl = LSD, D8 = MSD
(If the display reading is 1234, then D8 to D1 is: 00001234)

D9

Decimal Point (DP), position from right to the left

D10

Polarity: 0 = Positive 1 = Negative

D11 & D12 Annunciator for Display

°C = 01 Knot = 90 mile/h = 12

°F = 02 Km/h = 10 CMM = 84

m/S = 08 ft/min = 11 CFM = 85

D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0

Each digit indicates the following status:

Meter (3.5 mm jack plug) PC (9W ‘D” Connector)
Center Pin .................................. Pin 4 Pin 2 2.2 K
Ground/shield ............................ Pin 2 Pin 5 resistor

Block 1

Block 2

Block 250

………

………

Data 1 Data X+1 Data Z
To Data X To Data Y To Data 16,000

• Use the ▲UP (5) and ▼ DOWN (6) buttons to select the desired memory

block number (1 to 250) and push the SEND (7) button once.

• The bottom right of the display shows "Sending Data!".
• When the process has finished, "Transmit mode" is displayed.

• Push the ESC (3) button to exit this function.


6

. RS232 PC SERIAL INTERFACE


The instrument features a 3.5 mm RS232 OUTPUT TERMINAL (14). The

signal output is a 16-digit data stream that can be adapted to user-defined ap-
plications. A RS232 lead with the following connection is required to link the
instrument with the PC serial interface.