Calibration Procedure - Single Port
Check that the “CAL OFF” message is displayed in the upper left corner of the display.
This indicates that the Site Master has not been calibrated. The following procedure details
how to perform the calibration.
Step 1. Select the appropriate frequency range, as described in the procedure on page
Step 2. Press the
START CAL
key. The message
CONNECT OPEN TO RF Out
PORT will appear in the display.
Step 3. Connect the calibrated Open and press the
ENTER
key. The messages
Mea-
suring OPEN and then CONNECT SHORT TO RF Out PORT will appear.
Step 4. Remove the Open, connect the calibrated Short and press the
ENTER
key. The
messages
Measuring SHORT and then CONNECT TERMINATION TO RF
Out will appear.
Step 5. Verify that the calibration has been properly performed by checking that the
CAL ON message is now displayed in the upper left corner of the display.