CAUTION
This product has been evaluated with respect to
electrical shock, fire, and mechanical hazards
only in accordance with UL60601-1; IEC60601-1;
CAN/CSA C22.2 No. 601.1
WARNING
When performing a cauterization or similar treat-
ment, the patient must be insulated from the metal
portions of the table by nonconductive material.
Failure to do so may result in electrical shock or
burns to the patient.
WARNING
Use 120 VAC, 60 Hz alternating current only. Failure
to do so may result in electrical shock to personnel
and will result in damage to the table.
WARNING
Do not use this table in an explosive or oxygen-en-
riched atmosphere. Failure to do so may result in
serious personnel injury or death.
CAUTION
Do not use any power supply other than that
listed on the rating label (Figure 1). Failure to do
so may result in serious injury and/or equipment
damage.
SPECIFICATIONS
Electrical Requirements
... . . . . . . . . .120 VAC nominal, 60 Hz, 3 amps maximum
... . . . . . . . . . including maximum 2 amp accessory outlet
Fuse rating.................................3 Amp/250V/P/N 100091
Power Cord
................................ Extends 78 in. (Minimum) from table
................3 conductor, 120VAC. With hospital grade plug
Type of protection against electrical shock ........... Class 1
Type of protection
against electrical shock ............................ B applied parts
Type of protection against ingress of water……. Ordinary
Can accept paper rolls of.....................21 in. x 3.5 in. dia.
Certifications ..................................... UL Classified 60601
Pelvic Tilt ...............................................................5° Rise
Drawer Warmer ..................120 VAC, 65 watts, 0.5 amps
Weight of Table .....................................................220 lbs.
(With carton and skid)...........................................265 lbs.
Table Weight Capacity ..........................................500 lbs.
Stirrup Extension (maximum): ................................. 16 in.
Stirrup Lateral Positions ......................4 Lateral Positions
Leg Extension (maximum) ....................................... 15 in.
Dimensions:
Upholstered Top ....................28.0 in. wide x 57.3 in. long
With Leg Extension Fully
Extended .............................................................. 71.0 in.
Seat Height ...........32 in to top of upholstered seat at foot
Step .................. 7.2 in. high x 20.38 in. wide x 11 in. long
Step Capacity .......................................................500 lbs.
Base ............ 28.2 in. high x 23.25 in. wide x 46.0 in. long
Transportation and Storage
Temperature .............................................18°F to 176°F
Humidity......................................................-10% to 90%
Operation
Temperature .............................................50°F to 104°F
Humidity.......................................................30% to 75%
ELECTRICAL REQUIREMENTS
Electromagnetic Interference
This product is designed and built to minimize electromagnetic interference with other devices; however, if interfer-
ence is noticed between another device and this product, remove the interfering device from the room or plug this
product into an isolated circuit.